When Magnitude Suffices: Reconstructing Time-Domain-Gated Microwave Signals Without Phase on a Thoracic Phantom
Time-domain gating separates the arrivals that form a microwave S-parameter measurement. It requires the measurement of complex values, so this application almost always uses a vector network analyzer (VNA). We show that, on one calibrated measurement configuration, the gated magnitude the VNA would give can be reconstructed from a magnitude-only sweep of a spectrum analyzer with a tracking generator (SA). We select the gate center for separability of the measured states, correct each instrument once by a per-bin subtraction of a reference state, and train a fully connected network on sweeps from both instruments. The test object is a torso phantom with lung-shaped compartments whose fill...