Sign inGet started
Paper · arXiv

AEGIS: Assay-Aware Protocol Validation and Runtime Monitoring for Open-Source Liquid Handling Robots

Self-driving laboratories increasingly rely on low-cost liquid handlers such as the Opentrons OT-2, which ship without the pressure-based aspiration monitoring of Hamilton or Tecan systems and are typically run open-loop. Two failure modes go undetected: protocols that are syntactically valid but violate assay-specific invariants (e.g., tip reuse between a PCR template and a no-template control), and physical execution failures (partial dispense, air bubbles, missing tips) at runtime. We present AEGIS, a two-layer guardian for both. Layer 1 pairs a curated machine-readable assay rule database with an LLM that reasons over OT-2 Python code, reaching an adjusted F1 of 0.97 on a 24-protocol ...

AuthorsPriyanka V. Setty, Arvind Ramanathan, Ian Foster et al.
Primary categorycs.RO
Published17 July 2026
Last updated at source17 July 2026
cs.ROcs.AI
Embed this entry
<iframe src="https://silytics.si/embed/item/cfd66277-dd1a-4b01-abb0-8af0c36174c2" width="420" height="190" style="border:0" loading="lazy" title="Silytics entry"></iframe>

Free, no tracking. More about feeds and embeds.

Silytics shows public metadata from arXiv and links to the original. Information may be incomplete or out of date. Last refreshed 17 October 2026. Licenses and terms of the original apply to the work itself. Not financial, legal or professional advice.