Artifact Correction for Echo-Planar Imaging at Low-Field and Ultra-Low-Field MRI

Purpose: Echo-planar imaging (EPI) in low-field (LF) and ultra-low-field MRI (ULF) suffers from severe Nyquist ghost artifacts due to odd-even k-space misalignment. This study develops a reference-free artifact correction pipeline that reduces reliance on conventional reference scans while achieving improved ghost suppression. Methods: Starting from the traditional reference-scan-based ghost artifact correction method, we first introduce a peak-alignment-based ghost artifact correction method to correct odd-even line displacement without reference data. To further reduce residual artifacts, an interpolation-and-resampling strategy is applied. The combined method was evaluated using EPI an...